Sensitivity relations for electromagnetic flowmeters with rectangular and circular channels

L. M. Korsunskii

Abstract
A qualitative evaluation is presented for the sensitivity of a flow transducer of electromagnetic type, and ways of improving this are considered. It is shown that a device with a circular channel can provide a substantial improvement only if the velocity in the channel is increased by reducing the cross section. One can obtain an improvement by 1-2 orders of magnitude if one uses a rectangular channel and a design in which the walls of the channel are poles of the magnet.

Magnitnaya Gidrodinamika 8, No. 3, 124-130, 1972 [PDF] (in Russian)
Magnetohydrodynamics 8, No. 3, 390-394, 1972 [PDF, 0.27 Mb]

Copyright: Institute of Physics, University of Latvia
Electronic edition ISSN 1574-0579
Printed edition ISSN 0024-998X
DOI: http://doi.org/10.22364/mhd